Technical parameters
1. ★4.3-inch color touch screen and rotary encoder display operation;
2. ★ The instrument provides one set of pin plug-in electrode output, one set of
electro-acupuncture output, which are independently controllable and do not
interfere with each other;
3. Time setting function: the time range is adjustable from 0 to 99 min, and the single
step length is 1 min;
4. Output parameters of electrode therapy:
4.1 Each channel provides 1 set of electrode output, including two channels A and
B, and the output polarity of channel A and channel B is opposite.The output
waveform is a combination of square wave and exponential wave;
4.2 The pulse frequency is adjustable from 0.5 Hz to 10 Hz, when the frequency is
0.5 Hz to 1 Hz, the single step length is 0.1 Hz; when the frequency is 1 Hz to
10 Hz, the single step length is 1 Hz;
4.3 The pulse width is adjustable from 0.1ms to 10ms adjustable. When the pulse
width is 0.1ms~1ms, the single step length is 0.05ms; when the pulse width is
1ms~10ms, the single step length is 0.5ms;
4.4 Output intensity: current peak-to-peak value Ip-p is adjustable from 0mA to
99mA;
5. ★ The output parameters of electro-acupuncture treatment:
5.1 The carrier frequency is 500Hz ± 10%; the frequency of the modulation wave
is adjustable from 0.5Hz to 10Hz. When the frequency is 0.5 Hz to 1Hz, the
single step length is 0.1Hz; when the frequency is 1 Hz to 10Hz, the single step
length is 1Hz. The tolerance is ±10%;
5.2 The pulse width is adjustable from 0.1ms to 1ms, the single step length is
0.05ms, and the tolerance is ±10%;
5.3 Output intensity: Each channel of the electro-acupuncture of the therapeutic
apparatus is independently output. When the load impedance is 250Ω; the
peak-to-peak value Ip-p of each electroacupuncture output current is adjustable
from 0mA to 99mA, and the tolerance is ±15%.The effective value of the maximum
output current does not exceed 10mA;
7. Continuous working time is longer than 8h;
8. ★ The copyright of computer software for this product has been obtained;
9. Atmospheric pressure: 700hPa~1060hPa; rated voltage: ~220V±10%; power
frequency: 50Hz±1Hz;
Outer packing size(L*W*Hmm) |
Length(MM) |
Width (MM) |
Height(MM) |
Cubic meter |
N.W.(kg) |
packing
Weight |
G.W.(kg) |
Method of packing |
380*335*320 |
380 |
335 |
320 |
0.040736 |
4.1 |
0.5 |
4.6 |
Paper |