MY-C027C AED Defibrillator
Defibrillator
Category Specification
Output Biphasic Truncated exponential
Energy Sequence(child) 50J,50J,50J
Energy Sequence(Adult) 150J,200J,200J
Charge Time 6 sec, to 150J
8 sec. to 200 J(New battery)
Analysis Time 6 sec.
The maximum time from the initiation of rhythm analy-sis to readiness for discharge with a new battery. Less than 10seconds
The maximum time from the initiation of rhythm analy-sis to readiness for discharge after 6 shocks. Less than 10 seconds
The maximum time from initially switching power on toreadiness for discharge.Less than 18 seconds
Visual Prompts LED prompts;Automatic Self-test
Controls Four buttons: On/Off, Shock, Adult/child, CPR
Output Energy Accuracy ±10% into any impedance from 25 to 175Q
The Maximum Voltage 1400±50V
Output disabled when the patient impedance is outsidelimits 20Ω to 200Ω
Battery
Part No CR123A-4x2
Non-rechargeable 2.8 Ah
Capacity 120 discharges at 200 Joules
Shelf Life (25°C+15C) 5 years standby (after installation)
Physical
Dimensions 280 x 209 x 80 mm
Weight 2.5 kg
Operating Temperature 0°C to 40°ºC
Operating Humidity
Relative humidity between 30% and 95%(non-condensing)
Storage Temperature (without battery) -20°C to 55°ºC
Storage Humidity(without battery) Up to 93%(non-condensing)
Packing:Single package size: 33*16.5*36cm,weight 3.2kg 4 unit per carton, package size: 71*35*38.5cm, total gross weight 14 KG