| Electrical requirement |
~ 220V 50Hz or ~ 230V 60Hz |
| Power consumption |
≤ 450VA |
Incubator Mode
Mode of control: Air mode and Baby mode controlled by
micro-computer
Air temperature control range: 25°C - -37°C (override mode 37°C-39°C)
Baby temperature control range: 34°C- -37°C (overnde mode 37°C-38°C)
Accuracy of skin temperature sensor:±0.2°C
The temperature fluctuation between incubator temperature
and average incubator temperature (under stable temperature
condion):≤0.5°C
Temperature uniformity:≤0.8°C(mattress is on the horizontal position)
≤1.0°C(mattress is on the maximum tlt position)
Inclination of the bassinet: ±10° (stepless adjustable)
Noise inside hood (stable working condition):≤50dB(A)
Relative Humidity Level
Humidity display range: 0%RH-99%RH
Humidity control range: 0%RH~90%RH
Accuracy of humidity display:±5%RH
Accuracy of humidty control: ±5%RH
Oxygen Level
Indication range of oxygen concentration: 0%~99%
Control range of oxygen concentration: 20%-60%
Indication accuracy of oxygen concentration:
±2% (Oxygen concentration sttig value is 25% and below)
3% (Oxygen concentration stting value is 25% and above)
Control accuracy of oxygen concentration: ±4%