Technical parameters
1. Gross weight: 45kg size: 800mm * 780mm * 1150mm;
2. Net weight: 30kg size:700mm * 660mm * 950mm;
3. The adjustable range of cervical traction force: 0~300N. The step length is 1N. When the
traction force is adjusted to more than 200N, a warning will be issued and the operator will be
asked to confirm;
4. The average rate of change of traction force during the gradual and regressive phases of cervical
traction is 60N/s;
5. The equipment has a real-time monitoring function of traction, with a tolerance of ±30N;
6. Adjustable range of treatment time: 0~99min, step length is 1min;
7. Adjustable range of traction phase time: 0~9min, step length is 1min;
8. Adjustable range of intermittent phase time: 0~9min, step length is 1min;
9. The equipment has emergency protection measures. During the traction treatment, by pressing
the emergency button, the traction can be relaxed to the initial state;
10. The equipment has a neck heating belt, and the heating function can be turned on or off
separately.The maximum temperature does not exceed 41ºC;
11. The position corresponding to the front and back of the traction chair and the base of the
traction rope can be adjusted, and the adjustable range is 0-15 degrees;
12. The maximum weight of the patient that can be supported by the traction chair is 180kg;
13. Traction compensation: when the patient end is suddenly strained or relaxed due to external
force, the device should automatically restore the preset value;
14. Ambient temperature: 5ºC ~ 40ºC; Relative humidity: ≤80%; Atmospheric pressure:
860HPa ~ 1060HPa;
15. Power supply voltage: AC220V±10%, power supply frequency: 50Hz±2%; Input power: ≤
180V;
16. Electric traction therapy device is to use the traction feedback control mode, the operator
input parameters, through the control system into a given signal, by comparing a given signal
and sensor feedback signal to control the motor running, after reducer reducer, torque through
the drive chain and the main shaft to the drum, intertwining of traction rope traction in the Customer-centric, honesty as the foundation, result-oriented
roller. In the traction process, the actual traction force changes are detected by the sensor,
which can be timely fed back to the control system to realize the longitudinal traction of the
patient's cervical spine in real time;
17. ★ This product has obtained the copyright of computer software;
Outer packing size(L*W*Hmm) |
Length(MM) |
Width (MM) |
Height(MM) |
Cubic meter |
N.W.(kg) |
packing
Weight |
G.W.(kg) |
Method of packing |
785*770*1150 |
785 |
770 |
1150 |
0.6951175 |
|
|
45kg |
|