• External Sequential Compression Device My-S039-a
  • External Sequential Compression Device My-S039-a
  • External Sequential Compression Device My-S039-a

External Sequential Compression Device My-S039-a

Certification: ISO13485
Specification: L*W*Hmm 2720*720*920
Samples:
US$ 1205/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Diamond Member Since 2021

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company
  • Overview
  • Product Parameters
  • Detailed Photos
Overview

Basic Info.

Model NO.
MY-S039-A

Product Description

Product Parameters
Technical parameters:
1. Normal working conditions:
1.1. Working environment: temperature 5ºC~40ºC;
1.2. Relative humidity: 10%~80%;
1.3. Atmospheric pressure: 86kPa~106kPa;
1.4. External power supply: working voltage: ~ 220V±10%; power frequency: 50Hz
±1Hz;
1.5. Input power: 60VA.
2. Rotary knob and LCD touch screen , easy to operate;
3.With step pressure setting function
4. Time setting: continuous operation or setting function time range 0~99min, each
step 1min;
5. The adjustment range of the inflation interval time is 0-90seconds,each step
5seconds;
6. The pressure holding time is adjustable from 0-12seconds;
7.Treatment mode: 4 inflatable modes;
Mode A: Distal-to-proximal one-by-one (chamber) conventional inflation mode
Mode B: Progressive (chamber-by-chamber) inflation pattern from distal to proximal
Mode C: Progressive (chamber-by-chamber) inflation pattern from proximal to distal
Mode D: Regular inflation pattern by (chamber) from proximal to distal
8. Pressure indication: The therapeutic apparatus has a pressure indication to indicate
the therapeutic pressure generated by the therapeutic apparatus in the airbag
under the current treatment procedure;
9. Pressure adjustment: the airbag pressure adjustment range is 0~200 mmHg, and
the set adjustment step is 5 mmHg each;
10. The ultimate pressure is less than or equal to 300 mmHg, and the duration of
exceeding 15 mmHg is not more than 3 minutes;
11. Overpressure protection: The therapy device has an overpressure protection function;
12. Emergency stop switch: The therapy device provides an individual emergency
stop switch in addition to the power supply and function switch, which can stop
the treatment program at any time;
13. Manual pressure release: The therapy device provides manual release pressure
function for the patient under various conditions;
14. Air tightness: the airbag and the connecting pipeline should have good air
tightness, keep it for 1 min under the maximum output pressure, and the pressure
drop should not be more than 10%;
15. Connection: The connecting pipeline should have a device or mark to prevent
wrong connection;
16. The therapy instrument has the function of setting therapy parameters, which
can select the therapy mode, and airbag type, set the therapy pressure and therapy
time of each chamber;
17. The therapy device has a therapy parameter display function, which can display
the currently set therapy parameters, as well as the current inflation pressure,
remaining therapy time;
18. Working noise: The noise of the therapeutic apparatus during normal operation
should not be greater than 45dB (A)